Common failures of pressure testing machine 1:
The oil pump does deliver oil
Reason: There is air in the oil pump, the oil filter is blocked the oil outlet valve seat does match, the steel ball ball seat have scratches burrs.
Cleaning method: Open the high pressure oil outlet pipe joint of the oil pump to exhaust. Clean discharge the air in the oil pump. Replace repair corresponding parts.
Common failures of pressure testing machine 2:
The oil pump delivery is unstable (the pointer can see stagnation reciprocating vibration).
Reason: oil viscosity is too small (too thin) too dirty; there is air in the oil circuit. There is dirt roughness between the oil delivery valve piston its bushing. There is an oil leak.
Cleaning method: Replace clean oil with suitable viscosity. Sweep out the air in the oil circuit, after the piston rises for a certain distance, open the oil return plug. Pressure testing machine manufacturer cleaning: Grinding of napped parts. Find out the oil leak clean it up.
Common failures of pressure testing machine 3:
Oil pressure pulsation (the oil return pipe of the oil delivery valve appears intermittently, the dynamometer pointer of the standard dynamometer trembles when the load display value is checked)
Reason: There is air in the oil pump. The oil viscosity is too small (too thin). The oil delivery valve saves too much needle clearance. There is dirt in the oil pump. The oil outlet valve seat does match, the steel ball ball seat have burrs scratches.
Cleaning method: sweep the air. Replace with suitable oil. Cut to save needle gap. Clean the oil pump. Replace repair the corresponding parts.
Common failures of pressure testing machine 4:
Oil pressure can hit the maximum load
Reason: The oil plug of the oil delivery valve its sleeve are too tight there is dirt. The elastic force of the delivery valve spring is too small. There is a big oil leak. Oil is leaking the front end of the oil delivery valve piston. The oil pipe joint is leaking. The working piston clearance is too large. The needle valve of the oil return valve does match.
Cleaning method; cleaning grinding the corresponding parts. Add washer to the end of the spring replace the spring. Pull pressure testing machine manufacturers eliminate oil leakage. Tighten the screw sleeve. Tighten after replacing the gasket. Increase the oil viscosity slightly replace the piston. Combine the oil return valve.
Common failures of pressure testing machine 5:
The pointer does return to zero after unloading
Reason: The rack is stuck. The pendulum rod fell back too fast to make the rack jump out of the wheel.
Cleaning method: adjust the tension reed. Re-engage the rack. Rotate the rack to zero.